Resource Type

Journal Article 5

Year

2012 5

Keywords

BNCT 2

GBM 1

IHNI 1

PGNAA 1

absorbed dose rate of neutron and γ ray 1

boron concentration 1

boron concentration detect 1

boron neutron capture therapy(BNCT) 1

clinical trial 1

in hospital neutron irradiator 1

in-hospital neutron irradiatior for BNCT 1

in-hospital neutron irradiator(IHNI) 1

inherent safety 1

malignant melanoma 1

metastatic liver cancer 1

neutron capture therapy 1

neutron enengy spectrum 1

neutron flux density 1

recurent head & neck cancer 1

shielding 1

open ︾

Search scope:

排序: Display mode:

Preliminary characterization of radiation fields of IHNI-1 for BNCT

Chen Jun,Li Chunjuan,Song Mingzhe,Li Wei,Yao Shunhe,Diao Lijun,Yao Yanling,Zhang Zizhu,Gao Jijin,Zhang Yifan

Strategic Study of CAE 2012, Volume 14, Issue 8,   Pages 106-112

Abstract: neutron radiation fields produced by an in-hospital neutron irradiatior for boron neutron capture therapy(BNCT

Keywords: in-hospital neutron irradiatior for BNCT     neutron enengy spectrum     neutron flux density     absorbed dose    

The boron neutorn capture therapy (BNCT) status to go into the new era

Zhou Yongmao

Strategic Study of CAE 2012, Volume 14, Issue 8,   Pages 4-13

Abstract: : justify;">The paper briefly interprets the obvious progress of the boron neutorn capture therapy (BNCTIt includes the BNCT clinical positioning, the tumour recurrence exploring, the boron concentration quantifedThe enlargement of these bottle necks in BNCT developing might be the preview of personalizing and routineBNCT.

Keywords: neutron capture therapy     clinical trial     tumour recurrence     boron concentration detect     targeting boron compound     in hospital neutron irradiator    

Neutron capture therapy clinical development in the world

Zhang Zichu,Jin Congjun,Liu Kai,Zhang Guozhen,Yang Lijun

Strategic Study of CAE 2012, Volume 14, Issue 8,   Pages 100-105

Abstract:

Boron neutron capture therapy (BNCT) method was applied to about oneIn this paper, the principle of BNCT, the development history and international BNCT clinical progressThe BNCT clinical treatment situation and evaluation in glioblastoma (GBM), malignant melanoma, recurrent

Keywords: BNCT     GBM     malignant melanoma     recurent head & neck cancer     metastatic liver cancer    

PGNAA system preliminary design and measurement of IHNI

Zhang Zichu,Li Lihua,Gao Jijin,Fu Jinshu

Strategic Study of CAE 2012, Volume 14, Issue 8,   Pages 28-32

Abstract:

The preliminary design of PGNAA system of IHNI was completed and the system in accordance with the design was set up. For both detector shielding part and the neutron beam shielding part, the inner layer was natural LiF powder and the outer was lead.Then boron concentration measurement experiments using this system were finished.The results show that the the system can measure 10B concentration of 10 ppm, with 2 mL of the sample volume. 10B element measurement sensitivity was 0.822 cps/ppm. Also it was found that the natural LiF powder would produce a high background in the target part of the energy spectrum and affect the measurement accuracy.

Keywords: BNCT     IHNI     PGNAA     boron concentration     shielding    

Performance and characteristic of in-hospital neutron irradiator

Li Yiguo,Xia Pu,Peng Dan,Zou Shuyun,Wu Xiaobo,Zhang Jinhua,Zhang Zizhu,Gao Jijin,Wang Liyu,Fu Jinshu,Liu Tong,Zhou Yongmao

Strategic Study of CAE 2012, Volume 14, Issue 8,   Pages 20-22

Abstract:

The design of in hospital neutron irradiator (IHNI) and its systems were introduced, and the performance and characteristics of IHNI were described. In order to test the inherent safety of IHNI, the experiment of 4.2 mk reactivity release was done. The experimental results showed that the peak power of IHNI was 85.7 kW at the time of 229 s after 4.2 mk reactivity release, and then, the power decreased owing to the negative temperature coefficient of moderator. The radiation dose rates at different rooms were lower than the standard value.

Keywords: in-hospital neutron irradiator(IHNI)     boron neutron capture therapy(BNCT)     inherent safety    

Title Author Date Type Operation

Preliminary characterization of radiation fields of IHNI-1 for BNCT

Chen Jun,Li Chunjuan,Song Mingzhe,Li Wei,Yao Shunhe,Diao Lijun,Yao Yanling,Zhang Zizhu,Gao Jijin,Zhang Yifan

Journal Article

The boron neutorn capture therapy (BNCT) status to go into the new era

Zhou Yongmao

Journal Article

Neutron capture therapy clinical development in the world

Zhang Zichu,Jin Congjun,Liu Kai,Zhang Guozhen,Yang Lijun

Journal Article

PGNAA system preliminary design and measurement of IHNI

Zhang Zichu,Li Lihua,Gao Jijin,Fu Jinshu

Journal Article

Performance and characteristic of in-hospital neutron irradiator

Li Yiguo,Xia Pu,Peng Dan,Zou Shuyun,Wu Xiaobo,Zhang Jinhua,Zhang Zizhu,Gao Jijin,Wang Liyu,Fu Jinshu,Liu Tong,Zhou Yongmao

Journal Article