Microwave-Optical Frequency Standards at NIM:Realizing of Definition of Metre and Second

Li Tianchun

Engineering | CAE ›› 2007, Vol. 9 ›› Issue (6) : 27-31.

PDF(368 KB)
PDF(368 KB)
Engineering | CAE ›› 2007, Vol. 9 ›› Issue (6) : 27-31.

Microwave-Optical Frequency Standards at NIM:Realizing of Definition of Metre and Second

  • Li Tianchun

Author information +
History +

Abstract

  The paper presents the realization,  at NIM,  of SI time unit second using the NIM4 cesium fountain clock,  SI length unit meter using a femto-second (FS) optical frequency comb and calibration of wavelengths of stabilized lasers by this FS comb. The uncertainty of NIM4 fountain clock was evaluated as 5 × 10 - 15.  The frequency uncertainty of the FS optical comb locked to a H-maser, which is steered by NIM4 clock,  was estimated as 2. 2 × 10 - 14.  The function,  significance of Cs fountain clock, stabilized lasers,  optical comb and relationship among them are discussed. Finally the progress of NIM5 fountain clock and the construction of a Sr atom optical clock started in 2006 at NIM are reported.

Keywords

metrology / frequency standard / stabilized laser / Cs fountain clock / FS optical comb

Cite this article

Download citation ▾
Li Tianchun. Microwave-Optical Frequency Standards at NIM:Realizing of Definition of Metre and Second. Strategic Study of CAE, 2007, 9(6): 27‒31