Thermal/vacuum calibration of microwave humidity sounder on FY-3B satellite

Wang Zhenzhan,Zhang Shengwei,Li Jing,Li Yun,Wu Qiong

Strategic Study of CAE ›› 2013, Vol. 15 ›› Issue (10) : 33 -46.

PDF (2898KB)
Strategic Study of CAE ›› 2013, Vol. 15 ›› Issue (10) : 33 -46.

Thermal/vacuum calibration of microwave humidity sounder on FY-3B satellite

Author information +
History +
PDF (2898KB)

Abstract

The principle and methods on thermal/vacuum (T/V) testing of microwave humidity sounder (MWHS) on FY-3B satellite were described; correction algorithms on calibration errors were developed, and some main specifications of MWHS-3B, such as linearity, sensitivity and calibration precision, were obtained. The parameters derived from the T/V testing had been used in in-orbit data processing to produce brightness temperature products in FY-3B ground system.

Keywords

FY-3B satellite / microwave humidity sounder / prelaunch calibration / thermal/vacuum testing

Cite this article

Download citation ▾
Wang Zhenzhan,Zhang Shengwei,Li Jing,Li Yun,Wu Qiong. Thermal/vacuum calibration of microwave humidity sounder on FY-3B satellite. Strategic Study of CAE, 2013, 15(10): 33-46 DOI: