Journal Home Online First Current Issue Archive For Authors Journal Information 中文版

Engineering >> 2022, Volume 16, Issue 9 doi: 10.1016/j.eng.2020.10.021

Quadrupole-Linear Ion Trap Tandem Mass Spectrometry System for Clinical Biomarker Analysis

a Technology Innovation Center of Mass Spectrometry for State Market Regulation, Center for Advanced Measurement Science, National Institute of Metrology, Beijing 100029, China
b College of Instrument Science and Electrical Engineering, Jilin University, Changchun 130026, China

Received: 2020-07-05 Revised: 2020-09-19 Accepted: 2020-10-19 Available online: 2021-02-09

Next Previous

Abstract


The accurate and efficient measurement of small molecule disease markers for clinical diagnosis is of great importance. In this study, a quadrupole-linear ion trap (Q-LIT) tandem mass spectrometer was designed and built in our laboratory. Target precursor ions were first selected in the quadrupole, and then injected, trapped, and fragmented simultaneously in the linear ion trap (LIT) to reduce the space charge effect, enrich the target product ions, and promote sensitivity. The targeted analytes were measured with selected reaction monitoring using a positive scan mode with electrospray ionization (ESI). Ions with a mass-to-charge ratio (m/z) ranging from 195 to 2022 were demonstrated. When scanning at 1218 amu•s−1, unit resolution and an accuracy of higher than m/z 0.28 was obtained for m/z up to 2000. The dimensionless Mathieu parameter (q) value used in this study was 0.40 for collision-induced dissociation (CID), which was activated by resonance excitation. And an overall CID efficiency of 64% was achieved (activation time, 50 ms). Guanidinoacetic acid (GAA) and creatine (CRE) were used as model compounds for small molecule clinical biomarkers. The limits of quantification were 1.0 and 0.2 nmol•L−1 for GAA and CRE, respectively. A total of 77 actual samples were successfully analyzed by the home-built ESI-Q-LIT tandem mass spectrometry system. The developed method can reduce matrix interference, minimize space charge effects, and avoid the chromatographic separation of complex samples to simplify the pretreatment process. This novel Q-LIT system is expected to be a good candidate for the determination of biomarkers in clinical diagnosis and therapeutics.

SupplementaryMaterials

Figures

Fig. 1

Fig. 2

Fig. 3

Fig. 4

Fig. 5

Fig. 6

Fig. 7

Fig. 8

References

[ 1 ] Madhukar NS, Khade PK, Huang L, Gayvert K, Galletti G, Stogniew M, et al. A Bayesian machine learning approach for drug target identification using diverse data types. Nat Commun 2019;10():5221. link1

[ 2 ] Freeman JD, Rosman LM, Ratcliff JD, Strickland PT, Graham DR, Silbergeld EK. State of the science in dried blood spots. Clin Chem 2018;64(4):656–79. link1

[ 3 ] Cheillan D, Salomons GS, Acquaviva C, Boisson C, Roth P, Cordier MP, et al. Prenatal diagnosis of guanidinoacetate methyltransferase deficiency: increased guanidinoacetate concentrations in amniotic fluid. Clin Chem 2006;52(4):775–7. link1

[ 4 ] Mørkrid L, Rowe AD, Elgstoen KBP, Olesen JH, Ruijter G, Hall PL, et al. Continuous age- and sex-adjusted reference intervals of urinary markers for cerebral creatine deficiency syndromes: a novel approach to the definition of reference intervals. Clin Chem 2015;61(5):760–8. link1

[ 5 ] Benoit R, Samir M, Boutin J, Samuel A, Brigitte C, Dominique D, et al. LC–MS/MS measurements of urinary guanidinoacetic acid and creatine: method optimization by deleting derivatization step. Clin Chim Acta 2019;493:148–55. link1

[ 6 ] Tortorelli S, Eckerman JS, Orsini JJ, Stevens C, Hart J, Hall PL, et al. Moonlighting newborn screening markers: the incidental discovery of a second-tier test for Pompe disease. Genet Med 2018;20(8):840–6. link1

[ 7 ] Wagner M, Tonoli D, Varesio E, Hopfgartner G. The use of mass spectrometry to analyze dried blood spots. Mass Spectrom Rev 2016;35(3):361–438. link1

[ 8 ] Pang G, Chang Q, Bai R, Fan C, Zhang Z, Yan H, et al. Simultaneous screening of 733 pesticide residues in fruits and vegetables by a GC/LC–Q-TOFMS combination technique. Engineering 2020;6(4):432–41. link1

[ 9 ] Fernández-Fernández M, Rodríguez-González P, Añón Álvarez ME, Rodríguez F, Menéndez FVÁ, Alonso JIG. Simultaneous determination of creatinine and creatine in human serum by double-spike isotope dilution liquid chromatography–tandem mass spectrometry (LC–MS/MS) and gas chromatography–mass spectrometry (GC–MS). Anal Chem 2015;87(7): 3755–63. link1

[10] Smith CJ. Food safety and health—past problems and future solutions. Engineering 2020;6(4):384–8. link1

[11] Hou JJ, Zhang JQ, Yao CL, Bauer R, Khan IA, Wu WY, et al. Deeper chemical perceptions for better traditional Chinese medicine standards. Engineering 2019;5(1):83–97. link1

[12] Snyder DT, Fedick PW, Cooks RG. Multigenerational collision-induced dissociation for characterization of organic compounds. Anal Chem 2016;88(19): 9572–81. link1

[13] Cha B, Blades M, Douglas DJ. An interface with a linear quadrupole ion guide for an electrospray-ion trap mass spectrometer system. Anal Chem 2000;72(22): 5647–54. link1

[14] Snyder DT, Szalwinski LJ, Cooks RG. Simultaneous and sequential MS/MS scan combinations and permutations in a linear quadrupole ion trap. Anal Chem 2017;89(20):11053–60. link1

[15] Liu X, Wang X, Bu J, Zhou X, Ouyang Z. Tandem analysis by a dual-trap miniature mass spectrometer. Anal Chem 2019;91(2):1391–8. link1

[16] Guo D, Wang Y, Xiong X, Zhang H, Zhang X, Yuan T, et al. Space charge induced nonlinear effects in quadrupole ion traps. J Am Soc Mass Spectrom 2014;25(3): 498–508. link1

[17] Drogaris P, Le Blanc JCY, Fitzgerald JE, Lowndes NF, Verreault A, Thibault P. Enhanced protein detection using a trapping mode on a hybrid quadrupole linear ion trap (Q-trap). Anal Chem 2009;81(15):6300–9. link1

[18] Zhang X, Wang Y, Hu L, Guo D, Fang X, Zhou M, et al. Reducing space charge effects in a linear ion trap by rhombic ion excitation and ejection. J Am Soc Mass Spectrom 2016;27(7):1256–62. link1

[19] Hall AB, Coy SL, Kafle A, Glick J, Nazarov E, Vouros P. Extending the dynamic range of the ion trap by differential mobility filtration. J Am Soc Mass Spectrom 2013;24(9):1428–36. link1

[20] Xie J, Jiang H, Shen J, Peng T, Wang J, Yao K, et al. Design of multifunctional nanostructure for ultrafast extraction and purification of aflatoxins in foodstuffs. Anal Chem 2017;89(19):10556–64. link1

[21] Guna M, Londry FA. Tandem ion trap design with enhanced mass analysis capabilities for large populations of ions. Anal Chem 2011;83(16):6363–7. link1

[22] Douglas DJ, Frank AJ, Mao D. Linear ion traps in mass spectrometry. Mass Spectrom Rev 2005;24(1):1–29. link1

[23] Bier ME, Syka JEP, inventors; Finnigan Corporation, assignee. Ion trap mass spectrometer system and method. United States patent 5420425. 1995 May 30.

[24] Song Q, Kothari S, Senko MA, Schwartz JC, Amy JW, Stafford GC, et al. Rectilinear ion trap mass spectrometer with atmospheric pressure interface and electrospray ionization source. Anal Chem 2006;78(3):718–25. link1

[25] Perry RH, Cooks RG, Noll RJ. Orbitrap mass spectrometry: instrumentation, ion motion and applications. Mass Spectrom Rev 2008;27(6):661–99. link1

[26] Nolting D, Malek R, Makarov A. Ion traps in modern mass spectrometry. Mass Spectrom Rev 2019;38(2):150–68. link1

[27] Carducci C, Santagata S, Leuzzi V, Carducci C, Artiola C, Giovanniello T, et al. Quantitative determination of guanidinoacetate and creatine in dried blood spot by flow injection analysis-electrospray tandem mass spectrometry. Clin Chim Acta 2006;364(1–2):180–7. link1

[28] Gachumi G, Purves RW, Hopf C, El-Aneed A. Fast quantification without conventional chromatography, the growing power of mass spectrometry. Anal Chem 2020;92(13):8628–37. link1

[29] Nanita SC. Quantitative mass spectrometry independence from matrix effects and detector saturation achieved by flow injection analysis with real-time infinite dilution. Anal Chem 2013;85(24):11866–75. link1

[30] Sans M, Zhang J, Lin JQ, Feider CL, Giese N, Breen MT, et al. Performance of the MasSpec Pen for rapid diagnosis of ovarian cancer. Clin Chem 2019;65(5): 674–83. link1

[31] Chace DH. Mass spectrometry in newborn and metabolic screening: historical perspective and future directions. J Mass Spectrom 2009;44(2):163–70. link1

[32] Kvaskoff D, Heath AK, Simila HA, Ko P, English DR, Eyles DW. Minimizing matrix effects for the accurate quantification of 25-hydroxyvitamin D metabolites in dried blood spots by LC–MS/MS. Clin Chem 2016;62(4):639–46. link1

[33] Cognat S, Cheillan D, Piraud M, Roos B, Jakobs C, Vianey-Saban C. Determination of guanidinoacetate and creatine in urine and plasma by liquid chromatography–tandem mass spectrometry. Clin Chem 2004;50(8): 1459–61. link1

Related Research