Journal Home Online First Current Issue Archive For Authors Journal Information 中文版

Strategic Study of CAE >> 2012, Volume 14, Issue 8

PGNAA system preliminary design and measurement of IHNI

1. China Nuclear Institute Beijing 401 Hospital of China Institute of Atomic Energy, Beijing 102413, China;

2. Beijing Capture Technology Co. Ltd., Beijing 102413, China;

3. Inttadiation Metrology Division of China Institute of Atomic Energy, Beijing 102413, China

Received: 2012-05-26 Available online: 2012-07-26 09:05:08.000

Next Previous

Abstract

The preliminary design of PGNAA system of IHNI was completed and the system in accordance with the design was set up. For both detector shielding part and the neutron beam shielding part, the inner layer was natural LiF powder and the outer was lead.Then boron concentration measurement experiments using this system were finished.The results show that the the system can measure 10B concentration of 10 ppm, with 2 mL of the sample volume. 10B element measurement sensitivity was 0.822 cps/ppm. Also it was found that the natural LiF powder would produce a high background in the target part of the energy spectrum and affect the measurement accuracy.

Figures

图1

图2

图3

图4

图5

图6

图7

References

[ 1 ] Hatanaka H, Nakagawa Y. Clinical results of long-surviving brain tumo y patients who underwent boron neutron capture therapy[J]. Int J Radiat Oncol Biol Phys, 1994, 28 (5): 1061-1066. link1

[ 2 ] Kobayashi T, Kandak. Micoroanlysis systerm of PPM -order 10B concentration in Tissue for neutron capture therapy by prompt gam-ma -ray spectrometry [J]. Nuclear Instruments and Methods,1983, 204 (2-3):525-531. link1

[ 3 ] Harling O K, Chabeuf J M, Lambert F, et al. A prompt gamma neutron activation analysis facility using a diffracted beam [J].Nuclear Instruments and Methods in Phys Res -Section B, 1993,83(4):57-61. link1

[ 4 ] Riley K J, Harling O K. An improved prompt gamma activation analysis facility using a focused diffracted neutron beam [J]. Nucl Instr Meth, 143: 414-421. link1

[ 5 ] Crittin M, Kern J, Schenker J L. The new prompt gamma -ray activation facility at the Paul Scherrer Institute [J]. Switzerland Nuel. Instr. and Meth. A,2000, 449 (1-2):221-236. link1

[ 6 ] Byun S H, Choi H D. Design features of a prompt gamma neutron activation analysis system at HANARO [J]. Radioanal. Nucl.Chem, 2000, 244(2) :413-416. link1

[ 7 ] Byun S H, Sun G M, Choi H D. Development of a prompt gamma activation analysis facility using diffracted polychromatic neutron beam [J]. Nuel Instr and Meth, 2002, 487(3): 521-529. link1

[ 8 ] Byun S H, Sun G M, Choi H D. Beam characteristics of polychro-matie diffracted neutrons used for prompt gamma activation analysis[J]. Korean Nucl Soc,2002, 34 (1) :30-42. link1

Related Research